TT-270
  • The measuring methods of the TT-270 are magnetic induction (F) and eddy current (N). when a F series probe is connected, the unit measures non-magnetic coating on ferro substrates, when a N series probe is connected, the unit measures non-conductive coating on non-ferro substrates
  • Several types of probes are available for various applications:
    F400, F1, F1/90, F10, N1, CN02
  • Measurement modes: continuous / single
  • Automatic calculation: Mean values / Max. values / Min. values / No. of test, S.Dev.
  • Memory for maximum 640 readings
  • Working modes: direct mode (DIRECT) and Batch mode (APPL)
  • With backlight display
  • Integrated printer
  • Battery low indication
  • Switch off modes: manual and auto
Measuring range Refer to the table below
Probes available  F400, F1, F1/90, F10, N1, CN02
Tolerance Refer to the table below
Minimum resolution Refer to the table below
Measuring conditions
Refer to the table below
Operation language English
Standards DIN, ISO, ASTM, BS
Min. measuring area ø5mm (standard probe N1),
ø7mm (standard probe F1)
Calibration Zero and foil calibration
Statistics Maximum and minimum, mean,
standard deviation of 3000 readings,
number of measurements
Data memory 640 readings
Limits Min-max with alarm
Interface RS-232
Operating temperature -5°C - 40°C
Humidity
 20%~90%
Power supply NiMH rechargeable batteries, 1.25V
Dimensions 230mm x 86mm x 47mm
Weight 530gr





             
Probe model
F400 F1 F1/90º F10
N1 CN02
Operating principle Magnetic
induction
Magnetic
induction
Magnetic
induction
Magnetic
induction
Eddy current Eddy current
Measuring range (µm) 0-400 0-1250 0-10000 0 -1250
10-200
(for chromeplate on copper)
0-40
Low range resolution (µm) 0.1 0.1 0.1 10 0.1 1
Accuracy
One-point calibration (µm)
±(3%H+1) ±(3%H+1) ±(3%H+1) ±(3%H+10)
±(3%H+1.5) ±(3%H+1)
Accuracy
Two-point calibration (µm)
±[(1~3)H%+0.7] ±[(1~3)H%+1] ±[(1~3)H%+1] ±[(1~3)%H+10] ±[(1~3)%H+1.5] -
Measuring conditions
Min curvature of the min area (mm)
Convex 1
1.5
Flatten  
10 3 Flatten  
Min diameter of the area (mm) φ3 φ7 φ7 φ40 φ5 φ7
Critical thickness of substrate (mm) 0.2 0.5 0.5 2 0.3 unlimited
TT-270 F  Coating thickness gauge with integrated printer and F1 probe
TT-270 N  Coating thickness gauge with integrated printer and N1 probe
TT-270 FN  Coating thickness gauge with integrated printer and FN probe
  • Controler unit & integrated printer
  • Probe F1 or N1
  • Charger
  • Substrate & calibration foil set
  • Printer paper
  • Manual
  • INNOVATEST certificate
Related downloads
Optional accessories
  • Several probes for different applications